NES miHealth Recovery Guide / Factory Reset (Advanced)
For the NES miHealth (Gen 1)
Advanced procedure for the NES miHealth (gen 1 device) — only follow this if standard troubleshooting and a soft reset have not resolved the issue. This process requires BioSync software on a PC or Mac and involves putting the device into Recovery Mode. Contact customerservice@E4L.com before starting.
Steps:
Step 1 — Contact Customer Support
Contact customerservice@E4L.com and provide:
- Your miHealth Unit ID
- Your account email
How to find your Unit ID: Slide open the back battery panel — slide it gently towards the left to remove. You will see a sticker with a barcode. The second set of numbers and letters is your Unit ID.
Once support has your details, they will set up a recovery file on their end. You will receive confirmation before proceeding to the next step.
Step 2 — Connect Your miHealth in Recovery Mode
- Open BioSync and select miHealth Updater.
- Activate Recovery mode:
- first method: Press and hold the 2 up/down buttons on the left side of the miHealth as well as the round button below the screen. Press and release the power button tu turn on the device while keeping the other buttons pressed until the screen is white.
- alternative method: unplug the battery from the NES miHealth. Press and hold the round button below the screen and reconnect the miHealth battery, this should produce a white screen.
- With the white screen active, connect it to the computer using the cord provided by E4L. If your computer uses USB-C ports, use a block type adapter, not one with a cord. example: https://amzn.to/4b3Nwz7
- Once connected, BioSync will confirm the device is connected in Recovery Mode.
Step 3 — Recover Your miHealth
- Select Recover miHealth in BioSync.
- During the recovery, BioSync will appear to be loading or stuck and the miHealth will turn off and on several times.
- Set the miHealth down and do not touch it until you see a full blue screen with a "3x" and an arrow in the top left corner.
Important: Do not disconnect the device or close BioSync during this process, even if it appears frozen.
Step 4 — Calibration on the Device Screen
- Using the miHealth stylus (slide it out from the bottom of the right side of the miHealth) or the tip of a pen, gently tap the top left corner of the screen three times, then tap the bottom right corner three times.
- When prompted, lay the miHealth on a flat surface and tap the OK box on the screen.
- Place the miHealth on its right side and press OK again.
- The miHealth will restart automatically.
Step 5 — Post-Recovery Settings
- Disconnect the miHealth from the computer.
- Go to Settings (bottom right icon) and select Calibrate Electrodes — follow the on-screen prompts.
- Go back to Settings → Screen and choose the first option (the white dot will appear under the vertical miHealth icon).
- Return to the main screen.
- If the battery indicator is low, charge your miHealth before proceeding.
If the device still does not function after completing this recovery, contact customerservice@E4L.com for further assistance.