NES miHealth (Gen 1) — Key Limitations & Connection Requirements
For the NES miHealth (Gen 1)
⚠️ All BioSync connection and update instructions below apply to the Gen 1 (NES miHealth) only. If the user has a Gen 2 (E4L miHealth), BioSync is not compatible — direct them to the Wi-Fi update instructions in the "Updating & Syncing" section.
- No Wi-Fi or Bluetooth — the Gen 1 miHealth requires a physical cable connection to a Mac or Windows PC (not Chromebook)
- Updates and syncing via BioSync only — do not provide Gen 2 Wi-Fi update instructions to Gen 1 users
Cable types:
- The NES miHealth was produced in two versions — one with a USB-mini port and one with a USB-C port
- The original cable that came with the device must be used — while any cable will charge the miHealth, not all cables support data transfer
- Trying a different cable is the first troubleshooting step when a user has issues scanning or updating their NES miHealth
- If the original cable is lost, the user needs to identify which port their device has (USB-mini or USB-C) and source a compatible data transfer cable
⚠️ Do NOT recommend a Factory reset when BioSync cannot detect the device. The NES miHealth factory reset requires an active data connection through BioSync to complete. If performed without that connection, the device will enter white screen recovery mode and the user will be unable to use or recover their device. When cable and port troubleshooting fails, the correct escalation is to contact customerservice@e4l.com to have a force update enabled on the backend — not to attempt a device reset.
NES miHealth (Gen 1) — Troubleshooting: Soft Reset
This is the first troubleshooting step for any Gen 1 NES miHealth issue, unless the device is already showing a white screen or a blue screen with a yellow key (which have their own specific procedures below).
When to perform a soft reset:
- Device is less responsive than usual
- Screen flickers
- Device turns off during a cycle
- General odd behavior
- Recommended as regular maintenance for daily users
How to perform a soft reset:
- Make sure the miHealth is turned off.
- Press and hold the two side buttons (the intensity buttons on the left side of the device).
- While holding the side buttons, press the on/off button to turn the device on.
- Keep the side buttons pressed until the screen turns on.
NES miHealth (Gen 1) — How to Download Custom Folders
Custom folders (also called custom cycles) are downloaded to the Gen 1 miHealth via BioSync on a Mac or Windows PC.
Before you begin: The custom folder must first be created by the practitioner on the Recommendations screen inside a scan in the portal, and assigned to the correct miHealth device at portal.E4L.com.
- Turn on your miHealth and connect it to the computer using the original cord or one purchased from E4L.
- Open BioSync — check that the version number in the bottom-right says version 6.(we recommedn updating BioSync to the latest version for optimal results, but this is not necessary for the next step)
- Select miHealth Updater on the BioSync main screen.
- BioSync will refresh and display an image of a miHealth. Click Check for Updates in the bottom-right of the screen.
- If updates or custom folders are found, select Install Updates.
- The miHealth will download the custom folder. Do not touch the miHealth or BioSync until the process is complete. The miHealth will restart automatically when done.
- The custom folder will be visible on the main miHealth screen once the restart is complete.
NES miHealth (Gen 1) — How to Remove Custom Cycles
Custom cycles are removed via BioSync on a Mac or Windows PC.
- Turn on your miHealth and connect it to the computer.
- Open BioSync — check that the version number in the bottom-right says version 6 (we recommedn updating BioSync to the latest version for optimal results, but this is not necessary for the next step).
- Select miHealth Updater.
- Click Custom Cycles in the bottom-left (wait a few seconds for the cycles to load).
- Click the red/white cross button next to the cycle you wish to delete.
- Confirm that you want to delete the cycle.
- Click OK to close the confirmation popup.
- Repeat steps 5–7 for any other cycles you wish to delete.
- Click Close in the bottom-right to return to the miHealth Updater screen.
NES miHealth (Gen 1) — Troubleshooting: Device Restarts During a Program
Likely cause: Low or degraded battery.
Troubleshooting steps in order:
- Perform a soft reset (see Soft Reset instructions above).
- Fully charge the device and try again.
- If the battery is more than a year old and the soft reset does not resolve the issue, the battery likely needs replacing.
Replacement battery:
- Batteries are specific to the NES miHealth and cannot be purchased elsewhere.
- Order from the E4L portal shop: E4L Portal — Accessories
NES miHealth (Gen 1) — Troubleshooting: Screen Calibration
If screen presses become less accurate or the touchscreen is unresponsive, the screen can be re-calibrated. This process is similar to the soft reset but adds one extra button.
Buttons used: On/off button + two side buttons (intensity buttons) + top right button on the face of the device.
How to calibrate the screen:
- Make sure the miHealth is turned off.
- Press all four buttons at the same time and hold for 2 seconds.
- Release the on/off button and the two side buttons.
- Hold the top right button until the screen turns on, then release.
- If successful, you will see "3x" in the top left corner of the screen.
- Use the stylus to tap the top left corner three times, then do the same in the bottom right corner.
Watch the screen calibration video
NES miHealth (Gen 1) — Troubleshooting: Red X on Date & Time Screen
What it means: A red X will appear at the bottom of the date and time screen when the date/time on the device is set more than 13 hours from the current date/time.
The 13-hour limit exists to allow users to adjust their device when travelling across time zones, while preventing the date from being rolled back to extend Custom Cycles (which always have an expiry date).
Cause and resolution:
| Situation | Action |
|---|---|
| User has set the date/time more than 13 hours out | Correct the date and time to within 13 hours of the current local time |
| Device is restarting and setting itself to a date 30 years in the future | Contact E4L support and request a force update on the device |
| The 30-year future date issue is recurring | Contact support and flag it — the device may require a physical repair |
NES miHealth (Gen 1) — Troubleshooting: Blue Screen with Yellow Key
Cause: If the battery charge on the NES miHealth was very low several times in a row, and/or the battery was replaced, the miHealth may lock and display a blue screen with a yellow key. This requires a forced update.
A forced update is performed using BioSync software and a computer. The forced update will refresh your miHealth operating system and will clear the key symbol on your screen.
The forced update needs to be set by the E4L support team first. To do so please open a support ticket by visiting https://help.E4L.com/ scrolling to the bottom of the page and selecting Submit a Request
You will need to ask for a Forced Update and provide your E4L user email and the Unit id of your miHealth. The unit ID consists of numbers and letters. It can be found on the settings screen by selecting miUnit. It is also displayed on the sticker inside the battery compartment of your device under the barcode (second series of numbers and letters).
Once you receive a positive response from the support team, you can proceed. Open BioSync, connect your miHealth, turn your miHealth on, select miHealth updater on BioSync, and proceed with the update.